Related Experiment Video
Updated: Aug 1, 2025

05:34
Author Spotlight: Advancing Pathogen Diagnostics with Standardized LAMP
Published on: September 8, 2023
830
Commercially available SARS-CoV-2 RT-qPCR diagnostic tests need obligatory internal validation
Katarzyna Linkowska1, Tomasz Bogiel2, Katarzyna Lamperska3
1Department of Forensic Medicine, Collegium Medicum of the Nicolaus Copernicus University, Curie-Skłodowskiej Str. 9, 85-094, Bydgoszcz, Poland. linkowska@cm.umk.pl.
Scientific Reports
|April 28, 2023
Summary
Three SARS-CoV-2 diagnostic tests showed 100% specificity for the Omicron variant. Internal validation of limit of detection and cutoff Ct values is recommended over manufacturer claims for reliable COVID-19 variant detection.
Area of Science:
- Virology
- Molecular Diagnostics
- Public Health
Background:
- The ongoing evolution of SARS-CoV-2, particularly Omicron descendants, necessitates continuous vigilance and reliable diagnostic methods.
- Healthcare systems face persistent challenges due to increased transmissibility of new variants, making accurate detection imperative.
Purpose of the Study:
- To evaluate and compare the analytical performance of three commercially available SARS-CoV-2 diagnostic assays.
- To determine the specificity and sensitivity (Limit of Detection - LOD) of Liferiver, Vitaassay, and TaqPath RT-PCR kits for SARS-CoV-2, including the Omicron variant.
Main Methods:
- Analytical sensitivity and specificity were assessed using synthetic nucleic acid standards and clinical samples.
- Limit of Detection (LOD) was determined for each assay.
- Ct values were analyzed at specific viral RNA concentrations and compared to manufacturer-declared cutoff values.
Main Results:
- All three assays demonstrated 100% specificity for SARS-CoV-2 RNA and the Omicron variant.
- The LOD was 10 viral RNA copies/reaction for Liferiver and TaqPath, and 100 viral RNA copies/reaction for Vitaassay.
- Mean Ct values obtained were significantly lower than manufacturer-declared cutoffs, suggesting potentially higher sensitivity upon internal validation.
Conclusions:
- The evaluated RT-PCR assays are highly specific for SARS-CoV-2 and its Omicron variant.
- Internal validation of LOD and cutoff Ct values is crucial for accurate performance assessment, potentially revealing superior sensitivity compared to manufacturer specifications.
- Consistent and reliable detection of emerging SARS-CoV-2 variants remains critical for public health surveillance and management.