Monitoring the Wall Mechanics During Stent Deployment in a Vessel
Published on: May 8, 2012
Stent underexpansion is associated with high wall shear stress: a biomechanical analysis of the shear stent study
Sonali Kumar1, David Molony2, Sameer Khawaja1
1Division of Cardiology, Department of Medicine, Emory University School of Medicine, Atlanta, GA, USA.
Insights
Stent underexpansion in coronary arteries significantly increases the risk of high wall shear stress (WSS), a factor linked to adverse cardiovascular events. This finding highlights the importance of optimal stent deployment to prevent complications.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Interventional Cardiology
Background:
- Coronary stent underexpansion is a known risk factor for restenosis and stent thrombosis.
- High wall shear stress (WSS) in atherosclerosis is linked to prothrombotic pathways, matrix metalloproteinase activation, and myocardial infarction.
Purpose of the Study:
- To investigate the hypothesis that stent underexpansion predicts high WSS.
- To analyze the relationship between stent underexpansion and WSS distribution in angulated coronary arteries.
Main Methods:
- Prospective randomized controlled study comparing R-ZES and X-EES stents.
- Computational fluid dynamics (CFD) used with 3D reconstructions from intravascular ultrasound (IVUS) and angiography to calculate WSS.
- Logistic regression analysis to assess the association between WSS and underexpansion.
Main Results:
- Frames with stent underexpansion were more than twice as likely to exhibit high WSS (> 2.5 Pa) compared to non-underexpanded frames (OR = 2.197, p = 0.008).
- No significant association was found between underexpansion and low WSS (< 1.0 Pa).
- No significant differences in underexpansion were observed between the R-ZES and X-EES stent platforms.
Conclusions:
- Stent underexpansion is a significant predictor of high WSS in coronary arteries.
- These findings underscore the clinical importance of achieving optimal stent expansion to mitigate risks associated with high WSS.
- Further research may explore strategies to minimize stent underexpansion during percutaneous coronary intervention.
Abstract:
Coronary stent underexpansion is associated with restenosis and stent thrombosis. In clinical studies of atherosclerosis, high wall shear stress (WSS) has been associated with activation of prothrombotic pathways, upregulation of matrix metalloproteinases, and future myocardial infarction. We hypothesized that stent underexpansion is predictive of high WSS. WSS distribution was investigated in patients enrolled in the prospective randomized controlled study of angulated coronary arteries randomized to undergo percutaneous coronary intervention with R-ZES or X-EES. WSS was calculated from 3D reconstructions of arteries from intravascular ultrasound (IVUS) and angiography using computational fluid dynamics. A logistic regression model investigated the relationship between WSS and underexpansion and the relationship between underexpansion and stent platform. Mean age was 63±11, 78% were male, 35% had diabetes, mean pre-stent angulation was 36.7°±14.7°. Underexpansion was assessed in 83 patients (6,181 IVUS frames). Frames with stent underexpansion were significantly more likely to exhibit high WSS (> 2.5 Pa) compared to those without underexpansion with an OR of 2.197 (95% CI = [1.233-3.913], p = 0.008). There was no significant association between underexpansion and low WSS (< 1.0 Pa) and no significant differences in underexpansion between R-ZES and X-EES. In the Shear Stent randomized controlled study, underexpanded IVUS frames were more than twice as likely to be associated with high WSS than frames without underexpansion.
More Related Videos
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
Related Concept Videos
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Shearing Stresses in a Beam: Problem Solving
Thin-Walled Hollow Shafts
Shearing Stress
The average shearing stress can be calculated by dividing the shear by the area of the cross-section.
Stress Concentrations
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....