Analytical performance evaluation of two automated urine chemistry analysers using two levels of commercially
J A Hokamp1, K Freeman2, K E Harr3
1Department of Veterinary Biosciences, College of Veterinary Medicine, The Ohio State University, Columbus, OH, USA.
Objective:
Evaluate two point-of-care urine chemistry analysers, VetScan SA and VetLab UA using assayed, bilevel (two concentrations) urine quality control material to determine if performance is acceptable for semiquantitative clinical urine chemistry analysis.
Materials And Methods:
Normal and abnormal urine quality control material sent to 23 veterinary practices was evaluated three times by each clinic on in-clinic automated urinalysis instruments. Accuracy, precision and clinical utility were evaluated.
Results:
Normal urine quality control material: Results for blood, glucose, ketones and bilirubin were 100% accurate and precise for both analysers, and pH values were accurately acidic to neutral. However, pH from VetScan SA had clinically significant negative bias. Abnormal urine quality control material: VetScan SA: blood, microalbumin and bilirubin were 100% accurate; glucose, ketones, and protein demonstrated ≤10% inaccuracy; pH demonstrated 34% inaccuracy. VetLab UA: blood, ketones and bilirubin were 100% accurate; glucose and protein demonstrated ≤10% inaccuracy; pH was 100% accurately neutral to alkaline.
Clinical Significance:
VetScan SA had marked negative pH bias versus VetLab UA resulting in clinically significant, overly acidic results. Specific gravity, nitrite, and leukocyte test pads should not be used. Both instruments had excellent performance in normal quality control material. While blood, glucose, protein and bilirubin are correctly identified as present in abnormal quality control material, exact concentrations cannot be interpreted due to imprecision. Only semiquantitative results, not numerical values implying quantification, should be reported from urine test strips.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Data Validation
Key parameters for method validation include:
Quality Assurance
Development of Analytical Methods
Urine Studies I: Urinalysis
Qualitative Analysis
There are two main approaches to qualitative analysis:...