Validity and reliability of integrated pressure level real-ear-to-coupler difference measurements
Related Experiment Video
Updated: Jul 31, 2025

The Measurement of Unsteady Surface Pressure Using a Remote Microphone Probe
Published on: December 3, 2016
Matthew Urichuk1,2, David Purcell1,2,3, Susan Scollie1,2,3
1School of Communication Sciences and Disorders, Western University, London, Ontario, Canada.
Objectives:
(1) To validate the measurement of foam-tip real-ear-to-coupler differences (wRECD) using an integrated pressure level (IPL) method and (2) to compare the reliability of this method to SPL-based measurement of the wRECD.
Design:
SPL-based wRECD and the proposed IPL wRECD measurement were completed bilaterally. Test-retest reliability of IPL wRECD was determined with full re-insertion into the ear canal and compared to published SPL wRECD test-retest data.
Study Sample:
22 adults with normal hearing and middle ear status were recruited.
Results:
Differences between SPL-based wRECD and IPL wRECD measurements were within 1.51 dB on average below 5000 Hz. At and above 5000 Hz, IPL wRECD exceeded SPL wRECDs by 6.11 dB on average. The average test-retest difference for IPL wRECD across all assessed frequencies was 0.75 dB with the greatest improvements in reliability found below 750 Hz and above 3000 Hz.
Conclusions:
IPL wRECD yielded improved estimates compared to SPL wRECD in high frequencies, where standing-wave interference is present. Independence from standing wave interference resulted in increased wRECD values above 4000 Hz using the IPL measurement paradigm. IPL wRECD is more reliable than SPL wRECD, does not require precise probe-microphone placement, and provides a wider valid wRECD bandwidth than SPL-based measurement.
Related Concept Videos
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pressure Gauges
Measurement of Blood Pressure
Differential Leveling
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive...
Uncertainty in Measurement: Reading Instruments