Related Experiment Video
Updated: Jul 31, 2025

Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
Quantification of spinal bone marrow fat fraction using three-material decomposition technique on dual-energy CT: A
Baojun Li1, Ning Hua1, Janelle Li1
1Department of Radiology, Boston University School of Medicine, Boston, USA.
Background:
Two-material decomposition is insufficient to quantify the fat fraction of spinal bone marrow, which is comprised of a mixture of bone minerals, water, and yellow marrow (fat).
Purpose:
To develop an accurate three-material decomposition-based bone marrow fat fraction ( ) quantification technique for dual-energy CT.
Methods:
Bone marrow edema phantoms containing trabecular bone minerals, water, and fat were constructed using fat fractions and bone mineral density values matching those expected in healthy and edematous bone, and scanned on a commercial dual-energy CT. Fat quantified by were compared to MRI-based fat fraction ( ) and conventional two-material-decomposition-based fat fraction ( ) to evaluate its accuracy and dependency on various bone mineral densities.
Results:
demonstrated an excellent correlation with (r = 0.97, R2 = 0.96) in the phantom, significantly more accurate than FF2MD when confounding bone minerals are present (50 mg/cm3 : r = 1.02, R2 = 0.95 vs. r = 0.65, R2 = 0.79 (p < 0.01); 100 mg/cm3 : r = 0.81, R2 = 0.47 vs. r = 0.21, R2 = 0.21 (p < 0.05)).
Conclusions: