Related Experiment Video
Updated: Jul 31, 2025

Syringe-injectable Mesh Electronics for Stable Chronic Rodent Electrophysiology
Published on: July 21, 2018
Woven EndoBridge Device Migration and Microsnare Retrieval Strategy: Single Institutional Case Reports with Technical
Brandon A Santhumayor1, Timothy G White1, Cassidy Werner1
1Department of Neurosurgery, North Shore University Hospital, Donald and Barbara Zucker School of Medicine at Hofstra/Northwell, Manhasset, NY, USA.
Abstract:
The Woven EndoBridge (WEB) (MicroVention/Terumo) device is a treatment option for wideneck bifurcation aneurysms. An uncommon adverse effect is WEB device migration. While certain bailout strategies for WEB recovery have been described, there is still a paucity of information on optimal strategies to maximize both short and long-term post-operative outcomes. We add 2 cases at our institution to the existing literature of WEBectomy in the setting of complicated intracranial aneurysm treatment. We discuss the long-term imaging outcomes with additional fluoroscopy video demonstrating our technique. Our findings reflect a clear benefit for the use of the Amplatz GooseneckTM microsnare (Medtronic) device as a means of WEB recovery, coupled with potential stent-assisted WEB embolization to remove the aneurysm from the parent circulation, while minimizing recurrence and thromboembolic complications.