Related Experiment Video
Updated: Jul 31, 2025

Fast Inspection of Quality of Indigo Naturalis by Multiple Light Scattering
Published on: August 18, 2023
Portable near-infrared and Raman spectroscopic devices as complementary tools for instantaneous quality control of
Jordi Cruz1, Putthiporn Khongkaew2,3, Judit Puig Bertotto4
1EUSS School of Engineering, Barcelona, Spain.
Introduction:
Process analytical technology (PAT) guidance is implemented in the quality assurance of phytocompounds to achieve the Industry 4.0 concept. Near-infrared (NIR) and Raman spectroscopies are feasible for rapid, reliable quantitative analysis through transparent packaging without removing the samples from their original containers. These instruments can serve PAT guidance.
Objective:
This study aimed to develop online portable NIR and Raman spectroscopic methods for quantifying total curcuminoids in turmeric samples through a plastic bag. The method mimicked an in-line measurement mode in PAT compared with placing samples into a glass vessel (at-line mode).
Materials And Methods:
Sixty-three curcuminoid standard-spiked samples were prepared. Then, 15 samples were randomly selected as fixed validation samples, and 40 of the 48 remaining samples were chosen as calibration set. The results obtained from the partial least square regression (PLSR) models constructed by using the spectra acquired from NIR and Raman were compared with the reference values from high-performance liquid chromatography (HPLC).
Results:
The optimum PLSR model of at-line Raman was achieved with three latent variables and a root mean square error of prediction (RMSEP) of 0.46. Meanwhile, the PLSR model of at-line NIR with one latent variable offered an RMSEP of 0.43. For the in-line mode, PLSR models created from Raman and NIR spectra had one latent variable with RMSEP of 0.49 and 0.42, respectively. The R2 values for prediction were 0.88-0.92.
Conclusion:
The models established from the spectra from portable NIR and Raman spectroscopic devices with the appropriate spectral pretreatments allowed the determination of total curcuminoid contents through plastic bag.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
IR Spectrometers
Applications of IR Spectroscopy: Overview
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
UV–Vis Spectrometers
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...