Related Experiment Video
Updated: Jul 24, 2026

Electrochemiluminescence Assays for Human Islet Autoantibodies
Published on: March 23, 2018
Exploration of suitable external quality assessment materials for serum C-peptide measurement
Yuhang Deng1, Chao Zhang2, Bingqiong Li3
1National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Beijing Hospital, National Center of Gerontology, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, P.R. China.
Objectives:
To find suitable external quality assessment (EQA) materials for serum C-peptide, we evaluated the commutability of five types of processed materials.
Methods:
Seventy-four individual serum samples and 12 processed samples including three EQA samples currently in use, frozen human serum pools (FHSP), and three other kinds of processed samples were prepared by dissolving WHO International Standard Reagent for C-peptide (WHO ISR 13/146) in three different matrixes: 0.05 % bovine serum albumin, fetal bovine serum and human serum pools. Samples were analyzed using the isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method and six widely used immunoassays. The commutabilities of processed materials were assessed according to the difference in bias approach recommended by the IFCC. And the short- and long-term stability of FHSP samples at different temperatures were also evaluated.
Results:
Out of the five kinds of processed materials, FHSP samples were commutable on most assays. In contrast, the EQA materials currently in use were only commutable on a few immunoassays. Additionally, processed materials derived from WHO ISR 13/146 were found to be un-commutable on over half of immunoassays. The FHSP samples could be stably stored at 4 and -20 °C for at least 16 days, and at -80 °C for at least 1 year, but at room temperature only for 12 h.
Conclusions:
With clarified commutability and stability information, the human serum pool samples along with the developed ID-LC-MS/MS method could be used in the EQA program to promote the comparability among laboratories for C-peptide measurement in China.
More Related Videos
07:14Detection and Quantification of Calcitonin Gene-Related Peptide CGRP in Human Plasma Using a Modified Enzyme-Linked Immunosorbent Assay
Published on: June 16, 2023
07:15Measurement of Cyclic Guanosine Monophosphate (cGMP) in Solid Tissues using Competitive Enzyme-Linked Immunosorbent Assay (ELISA)
Published on: July 3, 2025