Noise Produced by Neonatal Ventilators Inside and Outside of the Incubators
Lea Bergez1, Gilles Jourdain1, Daniele De Luca2
1Division of Pediatrics and Neonatal Critical Care, "A.Beclere" Medical Center, Paris Saclay University Hospitals, APHP, Paris, France.
Background:
Insufficient data are available about the noise produced by modern neonatal ventilators. We aimed to measure their noise under different ventilatory modes and parameters.
Methods:
This was a bench study measuring the noise produced by 9 neonatal ventilators set in conventional or high-frequency oscillatory ventilation (HFOV), nasal mask-delivered CPAP with variable- or continuous-flow configuration, or bi-level positive airway pressure (considered as noninvasive ventilation [NIV]). Conventional ventilation and HFOV were tested in 2 distinct settings with moderate or higher parameters. Sound measurements were performed inside and outside an incubator mimicking the clinical setting and using a high-end meter meeting the international ISO 226:2003 standard.
Results:
Four ventilators remained below the internationally recommended safety threshold but only for measurements outside the incubator. Conventional ventilation (49.1 [3.4] dBA) and HFOV (56.3 [5.2] dBA) were the least and most noisy respiratory support technique, respectively. Noise was greater inside than outside the incubators (P < .0001) and different between the ventilators (P < .0001); better results were achieved by Servo-u and Fabian family devices for conventional ventilation; by fabian HFO for HFOV; and by Servo-u, VN500, and fabian family devices for CPAP and NIV. Noise levels were similar when using moderate or higher parameters in conventional ventilation (P = .81) and in HFOV (P = .45).
Conclusions:
Modern ventilators often produce relevant noise, independent of the respiratory support modality, with acceptable noise levels being measured only outside the incubator. Better results were achieved with Servo-u, VN500, and Fabian family devices.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...