Association between Intracanal Medicament Radiopacity and Streak Artifact Production Using Cone-beam Computed
Alizafar Wahidi1, William Nguyen Ha2, Lawrence Alvino1
1Adelaide Dental School, University of Adelaide, Adelaide, South Australia, Australia.
Introduction:
This study aimed to calculate the correlation between the radiopacity levels of various intracanal medicaments and radiolucent streak formation using cone-beam computed tomography (CBCT).
Methods:
Seven commercially-available intracanal medicaments were tested, which contained different amounts of radiopacifier [Consepsis, Ca(OH)2, UltraCal XS, Calmix, Odontopaste, Odontocide, and Diapex Plus]. Their radiopacity levels were measured according to the International Organization for Standardization 13116 testing standards (mmAl). Subsequently, the medicaments were placed in 3 canals of radiopaque artificial printed maxillary molars (n = 15 roots per medicament), leaving the second mesiobuccal canal empty. CBCT imaging was carried out using an Orthophos SL 3-dimensional scanner under recommended manufacturer exposure settings. Radiopaque streak formation was assessed by a calibrated examiner using a previously published grading system (0-3). The Kruskal-Wallis and Mann-Whitney U tests with and without Bonferroni correction were used to compare radiopacity levels and radiopaque streak scores for the medicaments. Their relationship was assessed using the Pearson correlation coefficient. The level of significance was set as (α = 0.05).
Results:
Diapex plus presented with the highest radiopacity levels (4.98 ± 0.01) and radiopaque streak scores [middle third (2.8 ± 0.18); apical third (2.73 ± 0.43)]; which was similar to UltraCal XS's radiopaque streak scores [middle third (2.8 ± 0.92); apical third (2.73 ± 0.77)]. Consepsis had the lowest radiopacity levels (0.12 ± 0.05), followed by Odontocide (0.60 ± 0.05). Consepsis and Ca(OH)2 were scored 0 for artifacts in all roots at all levels. A high positive correlation (R = 0.95) was found between radiopacity and streak formation.
Conclusions:
The radiopacity of intracanal medicaments varies and strongly correlates with the formation of radiolucent streak artifacts during CBCT.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
09:34Semi-Automated Planimetric Quantification of Dental Plaque Using an Intraoral Fluorescence Camera
Published on: January 27, 2023
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...