X-ray radiography of HPGe detectors for Monte Carlo simulations
Serhii Pohuliai1, Andrey Vlasenko1, Vladislav Malgin1
1Baltic Scientific Instruments, Ltd.(1), 3 Ramulu Street, Riga, LV-1005, Latvia.
Abstract:
The results of the development of a radiographic installation that allows to control the size of the internal elements of HPGe detectors' cryostats for simulation by Monte Carlo method, are presented. Calibration and assessment of image quality value of the developed radiographic installation were carried out in accordance with ISO Standards 19232-1 and 19232-5. Simulation of the HPGe detector's registration efficiency in the range of 50 keV-3 MeV with the MCC-MT program showed that the discrepancy between the calculated and experimental results when adjusting the detector model using radiological measurement data does not exceed 3% in geometry for point sources and 5% for the Eu-152 source in the Marinelli beaker geometry.