Related Experiment Video
Updated: Jul 31, 2025

Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Primary Stability of Collared and Collarless Cementless Femoral Stems - A Finite Element Analysis Study
Ryunosuke Watanabe1, Hajime Mishima1, Sho Totsuka1
1Department of Orthopaedic Surgery, Institute of Medicine, University of Tsukuba, Tsukuba, Japan.
Background:
Primary stability of the femoral stem is important for the long-term results of cementless total hip arthroplasty. Cementless collared stems have been known to have higher stability than collarless stems when there is a contact between the collar and the calcar. The purpose of this study was to compare the stabilities of collared stem and collarless stem in 2 loading conditions: 1) flat walking and 2) stair climbing.
Methods:
We constructed 3 finite element models. In the first model, the collar had contact with the calcar. The second model had a 1 mm gap between the calcar and the collar. The third model was constructed with a collarless stem. The proximal femur around the stem was divided into 3 zones: the upper zone (Gruen zones 1 and 7), the middle zone (Gruen zones 2 and 6), and the lower zone (Gruen zones 3 and 5). The micromotion at the stem/bone interface was measured at each zone of the 3 models under the 2 loading conditions.
Results:
The results showed that collared stems were more stable when the collar was in contact with the calcar than when a gap was left between the collar and the calcar. In particular, collar contact was highly effective in suppressing the micromotion proximal to the stem.
Conclusions:
Compared to the collarless stem, the collared stem had comparable stability when there was a gap at the collar and calcar interface and higher stability when there was contact between the collar and the calcar.
Related Concept Videos
Collar Bearings
Stability of structures
Bearings: Problem Solving
Design of Columns under a Centric Load
Euler's formula is applicable under the assumption that the column is a perfect, straight, homogenous prism, and it is operating...
Temperature Dependent Deformation
Euler's Formula for Pin-Ended Columns
To calculate the critical load,...