Related Experiment Video
Updated: Jul 31, 2025

Author Spotlight: Advancing Research on Candida albicans Biofilm-Associated Prosthetic Joint Infections
Published on: February 2, 2024
Indications and techniques for non-articulating spacers in massive bone loss following prosthetic knee joint
Jonas P Sina1, Shiraz A Sabah2, Daniel Schrednitzki3
1Nuffield Department of Orthopaedics, Rheumatology and Musculoskeletal Sciences (NDORMS), University of Oxford, Botnar Research Centre, Old Road, Oxford, OX3 7LD, UK. jonas.sina@ndorms.ox.ac.uk.
Introduction:
Prosthetic joint infection (PJI) is a destructive complication of knee replacement surgery (KR). In two-stage revision a spacer is required to maintain limb length and alignment and provide a stable limb on which to mobilise. Spacers may be articulating or static with the gold standard spacer yet to be defined. The aims of this scoping review were to summarise the types of static spacer used to treat PJI after KR, their indications for use and early complication rates.
Methods:
We conducted a scoping review based on the Joanna Briggs Institute's "JBI Manual for Evidence Synthesis" Scoping review reported following Preferred Reporting Items for Systematic Review and Meta-Analyses extension for Scoping Reviews (PRISMA-ScR) checklist. MEDLINE, EMBASE and CINAHL were searched from 2005 to 2022 for studies on the use of static spacers for PJI after KR.
Results:
41 studies (1230 patients/knees) were identified describing 42 static spacer constructs. Twenty-three (23/42 [54.2%]) incorporated cement augmented with metalwork, while nineteen (19/42, [45.9%]) were made of cement alone. Spacers were most frequently anchored in the diaphysis (22/42, [53.3%]), particularly in the setting of extensive bone loss (mean AORI Type = F3/T3; 11/15 studies 78.3% diaphyseal anchoring). 7.1% (79 of 1117 knees) of static spacers had a complication requiring further surgery prior to planned second stage with the most common complication being infection (86.1%).
Conclusions:
This study has summarised the large variety in static spacer constructs used for staged revision KR for PJI. Static spacers were associated with a high risk of complications and further work in this area is required to improve the quality of care in this vulnerable group.