Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Comparison of three calibration procedures for free-field reference speech audiometry
Mieke Janssens-Bij de Vaate1, Koenraad S Rhebergen1
1Department of Otorhinolaryngology and Head and Neck Surgery, Brain Center Rudolf Magnus, University Medical Center Utrecht, Utrecht, The Netherlands.
Three calibration methods for speech audiometry using TDH39 earphones were compared. All methods yielded similar results, indicating their utility for accurate speech recognition threshold testing.
Area of Science:
- Audiology
- Acoustics
- Speech Science
Background:
- IEC 60645-1 and ANSI S3.6 standards reference the free-field equivalent earphone output level for speech audiometry.
- Accurate calibration is crucial for reliable speech recognition threshold (SRT) measurements.
Purpose of the Study:
- To compare three distinct calibration procedures for the free-field equivalent earphone output method.
- To evaluate the effectiveness of these methods for TDH39 earphones in speech audiometry.
Main Methods:
- Speech audiometry was performed using Dutch consonant-vowel-consonant words under three conditions: TDH39 earphones, loudspeaker, and simulated free-field with TDH39 earphones.
- Three calibration procedures were assessed: empirical SRT difference, theoretical free-field correction, and simulated free-field results.
- Twenty normal-hearing subjects participated in the study.
Main Results:
- The observed difference between free-field and earphone SRT was 7.1 dB.
- The difference between free-field and simulated free-field conditions was 0.6 dB.
- All three calibration procedures produced comparable outcomes.
Conclusions:
- The three calibration procedures for the free-field equivalent output method are effective and interchangeable for TDH39 earphones.
- These findings support the use of these methods for accurate speech audiometry calibration.
More Related Videos
10:00Calibration of Vector Network Analyzer for Measurements in Radio Frequency Propagation Channels
Published on: June 2, 2020
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...