Comparison of three calibration procedures for free-field reference speech audiometry
Related Experiment Video
Updated: Jul 30, 2025

Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Mieke Janssens-Bij de Vaate1, Koenraad S Rhebergen1
1Department of Otorhinolaryngology and Head and Neck Surgery, Brain Center Rudolf Magnus, University Medical Center Utrecht, Utrecht, The Netherlands.
Objective:
In standards IEC 60645-1 and ANSI S3.6, the free-field equivalent earphone output level method is assumed as the reference for speech audiometry. Three calibration procedures for this method were compared in this study.
Design:
Speech audiometry was conducted with Dutch consonant-vowel-consonant words for the following conditions: 1. TDH39 earphones, 2. loudspeaker, and 3. free-field simulated with TDH39 earphones. The first calibration procedure was based on the empirically determined difference between the speech recognition threshold (SRT) with earphones and a loudspeaker. The second procedure was based on the theoretical free-field correction, derived from the known speech spectrum and the free-field to coupler difference. The third calibration procedure corresponded to the results of the free-field simulated speech material under earphones.
Study Sample:
The sample included 20 normal hearing subjects.
Results:
The differences between the observed SRT in the free-field and earphone conditions and the free-field and simulated free-field conditions were 7.1 dB and 0.6 dB, respectively.
Conclusion:
The three calibration procedures for the free-field equivalent output method yielded approximately the same results, and therefore all appear to be useful for TDH39 earphones.
More Related Videos
10:00Calibration of Vector Network Analyzer for Measurements in Radio Frequency Propagation Channels
Published on: June 2, 2020
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...