Related Experiment Video
Updated: Jul 30, 2025

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Ultrafast cadmium-zinc-telluride-based renal single-photon emission computed tomography: clinical validation
Matthieu Dietz1,2, Nicolas Jacquet-Francillon3, Alexandre Bani Sadr3
1Service de Médecine Nucléaire, Hospices Civils de Lyon, 59 Bvd Pinel, 69634, Lyon, France. matthieu.dietz@chu-lyon.fr.
Background:
One of the main limitations of 99mtechnetium-dimercaptosuccinic acid (DMSA) scan is the long acquisition time.
Objective:
To evaluate the feasibility of short DMSA scan acquisition times using a cadmium-zinc-telluride-based single-photon emission computed tomography (SPECT) system in children.
Materials And Methods:
The data of 27 children (median age: 4 years; 16 girls) who underwent DMSA SPECT were retrospectively analyzed. Both planar and SPECT DMSA were performed. SPECT images were analyzed using coronal-simulated planar two-dimensional images. A reduction in SPECT acquisition time was simulated to provide 4 series (SPECT-15 min, SPECT-10 min, SPECT-5 min and SPECT-2.5 min). A direct comparison of the planar and SPECT series was performed, including semi-quantification reproducibility, image quality (mean quality score on a scale of 0 to 2) and inter- and intra-observer reproducibility of the scintigraphic patterns.
Results:
The overall image quality score (± standard deviation) was 1.3 (± 0.6) for the planar data set, 1.6 (± 0.5) for the SPECT-15 min data set, 1.4 (± 0.5) for the SPECT-10 min data set, 1.0 (± 0.5) for the SPECT-5 min data set and 0.6 (± 0.6) for the SPECT-2.5 min data set. Median Kappa coefficients for inter-observer agreement between planar and SPECT images were greater than 0.83 for all series and all readers except one reader for the SPECT-2.5 min series (median Kappa coefficient = 0.77).
Conclusion:
Shortening SPECT acquisitions to 5 min is feasible with minimal impact on images in terms of quality and reproducibility.