Deep learning image reconstruction algorithms in low-dose radiation abdominal computed tomography: assessment of
Chun Yang1,2, Wenzhe Wang3, Dingye Cui1
1Department of Radiology, The First Affiliated Hospital of Shandong First Medical University & Shandong Provincial Qianfoshan Hospital, Jinan, China.
Background:
The image quality of computed tomography (CT) can be adversely affected by a low radiation dose, and reconstruction algorithms of an appropriate level may be useful in reducing this impact.
Methods:
Eight sets of CT images of a phantom were reconstructed with filtered back projection (FBP); adaptive statistical iterative reconstruction-Veo (ASiR-V) at 30% (AV-30), 50% (AV-50), 80% (AV-80), and 100% (AV-100); and deep learning image reconstruction (DLIR) at low (DL-L), medium (DL-M), and high (DL-H) levels. The noise power spectrum (NPS) and task transfer function (TTF) were measured. Thirty consecutive patients underwent low-dose radiation contrast-enhanced abdominal CT scans that were reconstructed using FBP, AV-30, AV-50, AV-80, and AV-100, and three levels of DLIR. The standard deviation (SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the hepatic parenchyma and paraspinal muscle were evaluated. Two radiologists assessed the subjective image quality and lesion diagnostic confidence using a 5-point Likert scale.
Results:
In the phantom study, both a higher DLIR and ASiR-V strength and a higher radiation dose led less noise. The NPS peak and average spatial frequency of the DLIR algorithms were closer to those of FBP, as the tube current increased and declined as the level of ASiR-V and DLIR strengthened. The NPS average spatial frequency of DL-L were higher than those of AISR-V. In clinical studies, AV-30 demonstrated a higher SD and lower SNR and CNR compared to DL-M and DL-H (P<0.05). For qualitative assessment, DL-M produced the highest qualitative image quality scores, with the exception of overall image noise (P<0.05). The NPS peak, average spatial frequency, and SD were the highest and the SNR, CNR, and subjective scores were the lowest with FBP.
Conclusions:
Compared with FBP and ASiR-V, DLIR provided better image quality and noise texture both in the phantom and clinical studies, and DL-M maintained the best image quality and lesion diagnostic confidence in low-dose radiation abdominal CT.
More Related Videos
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
05:41Author Spotlight: Integrating Ultrasound Imaging with Biochemical Markers for Thyroid Disease Diagnosis
Published on: February 9, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
X-ray Imaging