Impact of Fingernail Polish on Pulse Oximetry Measurements: A Systematic Review
Ashutosh Nath Aggarwal1, Ritesh Agarwal2, Sahajal Dhooria2
1Department of Pulmonary Medicine, Postgraduate Institute of Medical Education and Research, Chandigarh, India. aggarwal.ashutosh@outlook.com.
Background:
The effect of application of fingernail polish on SpO2 measurement remains unclear. We conducted this systematic review to ascertain the impact of fingernail polish on SpO2 measurement.
Methods:
We queried PubMed, Embase, and CINAHL databases for publications indexed through December 2022. We included studies providing paired SpO2 data from fingertips without and after nail polish application or reporting the number of subjects whose SpO2 could not be measured due to fingernail polish. We used random effects modeling to summarize standardized mean differences (SMDs) and corresponding 95% CI for different nail polish colors from comparative studies.
Results:
We retrieved 122 studies and included 21 publications, mostly performed on healthy volunteers. Of these, 17 (81.0%) studies had a low risk of bias. We summarized mean SMD for 10 nail polish colors (black, blue, brown, green, orange, pink, purple, red, white, and yellow) from 25 paired data sets on SpO2 across 20 studies. We found small (likely clinically insignificant) but statistically significant differences in mean SpO2 when fingers were coated with black, blue, brown, or purple nail polish (SMD -0.57, -0.47, -0.33, and -0.25, respectively; 95% CI -0.86 to -0.29, -0.84 to -0.10, -0.59 to -0.07, and -0.48 to -0.02, respectively). Only one of 4 studies reported a high proportion of unsuccessful oximeter readings from fingers painted with black (88.0%) or brown (36.0%) nail polish.
Conclusions:
Although fingernail polish of some colors can marginally reduce SpO2 reading or occasionally impede SpO2 measurement, the variability is clinically insignificant.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Pulse
The pulse serves as a clinical...