The influences of finite aperture size in photoacoustic computed tomography
Xiaofei Luo1, Jinsheng Jiang1, Hualin Wu2
1Department of Biomedical Engineering, School of Basic Medical Science, Central South University, Changsha, Hunan 410083, China.
Abstract:
In photoacoustic computed tomography (PACT), the "finite aperture effect" is often characterized as a tangential resolution that increases proportionally with the distance from the rotation center. However, this conclusion is based on the inaccurate point-detector assumption used in image reconstruction. In this study, we appropriately modeled the finite size of the acoustic detector in the back-projection (BP) based image reconstruction to improve the accuracy of the time delay calculation and systematically investigated its effects. Our results showed that the main effect of the finite aperture size is the creation of a limited high-quality imaging region (HQIR) around the scanning center, due to the directional sensitivity of the detector. We also demonstrated that the "finite aperture effect" can reduce the optimal number of detectors required for spatial anti-aliasing. These new findings provide novel perspectives for optimizing PACT systems and corresponding reconstruction methods.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...