DLCO Biologic Quality-Control Findings From a Multi-Center Global Study

Ellen A Becker1, Ellen M Moran2, Jason B Blonshine3

  • 1Rush University Medical Center, Chicago, Illinois. Ellen_Becker@rush.edu.

Respiratory Care
|May 16, 2023
PubMed
Summary

A diffusing capacity of the lung for carbon monoxide (DLCO) biologic quality control (BioQC) coefficient of variation (CV) of 6% is achievable. This ensures DLCO measurements are within an expected range, similar to ATS/ERS standards.