Evaluation and comparison of ocular biometric parameters obtained with Tomey OA-2000 in silicone oil-filled aphakic
Yongqun Xiong1, Yongdong Lin1, Zifeng Zhao1
1Joint Shantou International Eye Center of Shantou University, The Chinese University of Hong Kong, Dong xia Road, Shantou, Guangdong Province, People's Republic of China.
Purpose:
To evaluate a new non-contact instrument (OA-2000) measuring the ocular biometry parameters of silicone oil (SO)-filled aphakic eyes, as compared with IOLMaster 700.
Methods:
Forty SO-filled aphakic eyes of 40 patients were enrolled in this cross-sectional clinical trial. The axial length (AL), central corneal thickness (CCT), keratometry ((flattest keratometry) Kf and (steep keratometry, 90° apart from Kf) Ks), and axis of the Kf (Ax1) were measured with OA-2000 and IOLMaster 700. The coefficient of variation (CoV) was calculated to assess the repeatability. The correlation was evaluated by the Pearson coefficient. Bland-Altman analysis and paired t test were used to analyze the agreements and differences of parameters measured by the two devices, respectively.
Results:
The mean AL obtained with the OA-2000 was 23.57 ± 0.93 mm (range: 21.50 to 25.68 mm), and that obtained with the IOLMaster 700 was 23.69 ± 0.94 mm (range: 21.85 to 25.86 mm), resulting in a mean offset of 0.124 ± 0.125 mm (p < 0.001). The mean offset of CCT measured by OA-2000 and IOLMaster 700 was 14.6 ± 7.5 μm (p < 0.001). However, the Kf, Ks and Ax1 values from the two devices were comparable (p > 0.05). All the measured parameters of the two devices showed strong linear correlations (all r ≥ 0.966). The Bland-Altman analysis showed a narrow 95% limits of agreement (LoA) of Kf, Ks and AL, but 95%LoA of CCT and Ax1 was wide, which were - 29.3 ~ 0.1 μm and-25.9 ~ 30.7°respectively. The CoVs of the biometric parameters obtained with OA-2000 were lower than 1%.
Conclusion:
In SO-filled aphakic eyes, the ocular parameters (including AL, Kf, Ks, Ax1, and CCT) measured by the OA-2000 and IOLMaster 700 had a good correlation. Two devices had an excellent agreement on ocular biometric measurements of Kf, Ks and AL. The OA-2000 provided excellent repeatability of ocular parameters in SO-filled aphakic eyes.