Related Experiment Video
Updated: Jul 30, 2025

Cryo-Structured Illumination Microscopic Data Collection from Cryogenically Preserved Cells
Published on: May 28, 2021
Scanning SQUID-on-tip microscope in a top-loading cryogen-free dilution refrigerator
Haibiao Zhou1, Nadav Auerbach1, Indranil Roy1
1Department of Condensed Matter Physics, Weizmann Institute of Science, Rehovot 7610001, Israel.
Abstract:
The scanning superconducting quantum interference device (SQUID) fabricated on the tip of a sharp quartz pipette (SQUID-on-tip) has emerged as a versatile tool for the nanoscale imaging of magnetic, thermal, and transport properties of microscopic devices of quantum materials. We present the design and performance of a scanning SQUID-on-tip microscope in a top-loading probe of a cryogen-free dilution refrigerator. The microscope is enclosed in a custom-made vacuum-tight cell mounted at the bottom of the probe and is suspended by springs to suppress vibrations caused by the pulse tube cryocooler. Two capillaries allow for the in situ control of helium exchange gas pressure in the cell that is required for thermal imaging. A nanoscale heater is used to create local temperature gradients in the sample, which enables quantitative characterization of relative vibrations between the tip and the sample. The spectrum of the vibrations shows distinct resonant peaks with a maximal power density of about 27 nm/Hz1/2 in the in-plane direction. The performance of the SQUID-on-tip microscope is demonstrated by magnetic imaging of the MnBi2Te4 magnetic topological insulator, magnetization and current distribution imaging in a SrRuO3 ferromagnetic oxide thin film, and thermal imaging of dissipation in graphene.