Subcutaneous defibrillator use with unipolar pacemakers: Cautiously possible
Matanyahu Rubinstein1, Angela Zhu1, Justin A Mariani2
1Department of Cardiology, Heart Centre, Alfred Health, Melbourne, Australia.
Abstract:
Implantation of subcutaneous implantable cardioverter-defibrillator in a patient with existing unipolar pacemaker is against manufacturer recommendations. We report the case of a successful subcutaneous implantable cardioverter-defibrillator implantation in a patient with Fontan circulation and concomitant active unipolar pacing and present a summary of recommendations when considering subcutaneous implantable cardioverter-defibrillator implantation with unipolar pacing. Recommendations included: pre-procedure screening, rescreening during implantation and ventricular fibrillation induction, pacemaker programming, and post-procedure investigations.
More Related Videos
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Depolarizing Blockers: Pharmocokinetics
Cardiomyopathy V: Interprofessional Care