Related Experiment Video
Updated: Jul 30, 2025

Measurement of Tissue Oxygenation Using Near-Infrared Spectroscopy in Patients Undergoing Hemodialysis
Published on: October 2, 2020
Pilot Study of a Wearable Hydration Monitor in Haemodialysis Patients: Haemodialysis Outcomes & Patient Empowerment
Vicki Sandys1,2, Lavleen Bhat1, Emer O'Hare1
1Royal College of Surgeons in Ireland, Dublin, Ireland.
Introduction:
We aimed to assess the validity and reproducibility of a wearable hydration device in a cohort of maintenance dialysis patients.
Methods:
We conducted a prospective, single-arm observational study on 20 haemodialysis patients between January and June 2021 in a single centre. A prototype wearable infrared spectroscopy device, termed the Sixty device, was worn on the forearm during dialysis sessions and nocturnally. Bioimpedance measurements were performed 4 times using the body composition monitor (BCM) over 3 weeks. Measurements from the Sixty device were compared with the BCM overhydration index (litres) pre- and post-dialysis and with standard haemodialysis parameters.
Results:
12 out of 20 patients had useable data. Mean age was 52 ± 12.4 years. The overall accuracy for predicting pre-dialysis categories of fluid status using Sixty device was 0.55 [K = 0.00; 95% CI: -0.39-0.42]. The accuracy for the prediction of post-dialysis categories of volume status was low [accuracy = 0.34, K = 0.08; 95% CI: -0.13-0.3]. Sixty outputs at the start and end of dialysis were weakly correlated with pre- and post-dialysis weights (r = 0.27 and r = 0.27, respectively), as well as weight loss during dialysis (r = 0.31), but not ultrafiltration volume (r = 0.12). There was no difference between the change in Sixty readings overnight and the change in Sixty readings during dialysis (mean difference 0.09 ± 1.5 kg), [t(39) = 0.38, p = 0.71].
Conclusion:
A prototype wearable infrared spectroscopy device was unable to accurately assess changes in fluid status during or between dialysis sessions. In the future, hardware development and advances in photonics may enable the tracking of interdialytic fluid status.
More Related Videos
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
Related Concept Videos
Hemodialysis III: Nursing Management
Hemodialysis II: Procedure and Complications
Holter Monitor: 24-Hour Monitoring
Hemodialysis I: Introduction
Heart Failure VI: Adjunct Therapies
Peritoneal Dialysis III: Nursing Management