Masseter and digastric muscle activity evaluation using a novel electromyogram that utilizes elastic sheet electrodes
Akio Fusayama1, Tomoaki Mameno1, Masahiro Wada1
1Department of Prosthodontics, Gerodontology, and Oral Rehabilitation, Osaka University Graduate School of Dentistry, Suita, Japan.
Journal of Prosthodontic Research
|May 17, 2023
Summary
A new flexible electromyogram (EMG) device reliably measures muscle activity during mastication and swallowing. This novel device demonstrates high reproducibility and accuracy compared to conventional EMG systems.
Area of Science:
- Biomedical Engineering
- Neuroscience
- Kinesiology
Background:
- Accurate measurement of muscle activity is crucial for understanding mastication and swallowing.
- Conventional electromyogram (EMG) devices may have limitations in flexibility and comfort for these specific applications.
Purpose of the Study:
- To evaluate the reproducibility and reliability of a novel electromyogram (EMG) device utilizing a flexible sheet sensor.
- To assess the device's capability in measuring muscle activity related to mastication and swallowing.
Main Methods:
- Developed a new EMG device with elastic sheet electrodes for masseter and digastric muscle activity measurement.
- Assessed measurement reproducibility using intraclass correlation coefficient (ICC).
- Evaluated reliability by comparing key parameters (maximum amplitude, duration, integrated value, SNR) with conventional EMG devices using ICC and Bland-Altman analysis.
Main Results:
- The new EMG device exhibited high reproducibility with ICC scores of 0.92 (ICC (1,1)) and 0.88 (ICC (2,1)).
- High correlations were observed when compared to active electrode EMG devices for maximum amplitude (0.90), duration (0.99), integrated values (0.90), and SNR (0.75), with no significant fixed or proportional errors.
- Comparisons with passive electrode EMG devices showed high correlations for maximum amplitude (0.73) and duration (0.89), although SNR showed a significant fixed error.
Conclusions:
- The novel flexible sheet sensor EMG device demonstrates high reliability and reproducibility.
- This new EMG device is suitable for evaluating muscle activity during mastication and swallowing functions.