Precise evaluation of blood flow patterns in human carotid bifurcation based on high-frame-rate vector flow imaging

Shaofu Hong1, Yinghui Dong1, Di Song1

  • 1Department of Ultrasound, Shenzhen People's Hospital (The Second Clinical Medical College, Jinan University; The First Affiliated Hospital, Southern University of Science and Technology), Shenzhen, Guangdong, China.

Insights

High-frame-rate vector flow imaging (HiFR-VFI) reliably detects complex blood flow in the carotid bifurcation, outperforming traditional color Doppler. This advanced ultrasound technique offers quantitative turbulence measurement for assessing arterial disease.

Area of Science:

  • Medical Imaging
  • Cardiovascular Ultrasound
  • Hemodynamics

Background:

  • Carotid bifurcation (CB) assessment is crucial for cardiovascular health.
  • Traditional ultrasound methods like color Doppler flow imaging (CDFI) have limitations in precisely characterizing complex flow patterns.
  • High-frame-rate vector flow imaging (HiFR-VFI) offers potential for more detailed hemodynamic evaluation.

Purpose of the Study:

  • To compare the feasibility of HiFR-VFI against CDFI for evaluating flow characteristics in the carotid bifurcation of healthy adults.
  • To assess the accuracy and capabilities of HiFR-VFI in identifying and quantifying non-laminar and complex blood flow patterns.

Main Methods:

  • Forty-three healthy volunteers underwent carotid bifurcation flow assessment using both HiFR-VFI and CDFI.
  • Flow patterns were classified based on streamlines visualized by HiFR-VFI.
  • A novel turbulence index (Tur-value) was quantitatively measured, and interobserver agreement was evaluated.

Main Results:

  • HiFR-VFI identified non-laminar flow in 18.6% of cases where CDFI did not.
  • HiFR-VFI revealed a significantly larger extension of complex flow (0.37 cm² vs 0.22 cm²).
  • Quantitative analysis showed distinct Tur-values for different flow patterns (laminar, rotational, reversed, complex), with complex flow exhibiting the highest turbulence.

Conclusions:

  • HiFR-VFI reliably characterizes complex hemodynamics in the carotid bifurcation.
  • Quantitative turbulence measurement using HiFR-VFI provides valuable insights.
  • HiFR-VFI shows promise as an auxiliary diagnostic tool for atherosclerotic arterial disease assessment.