Assessment of Platelet Function by Automated Light Transmission Aggregometry
Mayukh K Sarkar1, Christine Hinz2
1Sysmex America Inc., Lincolnshire, IL, USA. sarkarm@sysmex.com.
Methods in Molecular Biology (Clifton, N.J.)
|May 19, 2023
Summary
Automated analyzers offer standardized platelet function testing, moving beyond manual light transmission aggregometry. Sysmex CS-Series and CN-Series analyzers provide new methods for routine laboratory platelet aggregation measurement.
Area of Science:
- Clinical diagnostics
- Hematology
- Hemostasis and thrombosis
Background:
- Light transmission aggregometry (LTA) is the established gold standard for platelet function testing.
- LTA is typically confined to specialized laboratories due to its manual and labor-intensive nature.
- Automated testing offers enhanced standardization and accessibility for platelet function analysis.
Purpose of the Study:
- To describe platelet aggregation measurement using Sysmex CS-Series and CN-Series routine blood coagulation analyzers.
- To outline the methodological differences between the CS-5100 and CN-6000 analyzers for platelet function testing.
Main Methods:
- Platelet aggregation was measured on the Sysmex CS-5100 and CN-6000 analyzers.
- CS-5100 involves manual pipetting for concentrated agonist preparation, with final dilution within the analyzer.
- CN-6000 utilizes an auto-dilution feature for automatic preparation of agonist dilutions and working concentrations.
Main Results:
- Both CS-Series and CN-Series analyzers enable platelet aggregation measurement.
- The CS-5100 requires manual preparation of concentrated agonists.
- The CN-6000 automates agonist dilution and preparation.
Conclusions:
- Sysmex CS-Series and CN-Series analyzers provide automated platforms for platelet function testing.
- These automated systems facilitate the performance of platelet aggregation tests in routine laboratory settings.
- Methodological differences exist in agonist preparation between the CS-5100 and CN-6000 analyzers.