Adverse events associated with AveirTM VR leadless pacemaker: A Food and Drug Administration MAUDE database study
Jalaj Garg1, Kuldeep Shah2, Rahul Bhardwaj1
1Cardiac Arrhythmia Service, Loma Linda University Health, Loma Linda, California, USA.
Background:
Leadless pacemaker (LP) offers an innovative approach for treating bradyarrhythmia, thus avoiding pacemaker pocket and lead-related complications. The Food and Drug Administration (FDA) has recently approved the Aveir™ leadless pacing system (screw-in type LP).
Methods:
We queried the FDA MAUDE database to study the safety profile and assess the types of complications with this relatively novel device technology. A MAUDE database search was conducted on January 20, 2023, for reports received post-FDA approval to capture all adverse events.
Results:
A total of 98 medical device report were reported for Aveir™ LP. After excluding duplicate, programmer-related, or introducer-sheath-related entries (n = 34), 64 entries were included. The most commonly encountered problem was high threshold/noncapture (28.1%, 18 events), followed by stretched helix (17.2%, 11 events) and device dislodgement (15.6%, ten events-5 intraprocedural, while 5 in the postoperative Day 1). Other reported events included high impedance (14.1%, nine events), sensing issues (12.5%, eight events), bent/broken helix (7.8%, five events), premature separation (4.7%, three events), interrogation problem (3.1%, two events), low impedance (3.1%, two events), premature battery depletion (1.6%, one event) and inadvertent MRI mode switch (1.6%, one event) and miscellaneous (15.6%, n = 10). There were eight serious patient injury events-pericardial effusion requiring pericardiocentesis (7.8%, five events) due to cardiac perforation that resulted in two deaths (3.1%) followed by sustained ventricular arrhythmias (4.6%, n = 3).
Conclusion:
In our study assessing the real-world safety profile of the Aveir™ LP, serious adverse events have been reported-life-threatening ventricular arrhythmias, pericardial effusion, device explantation/reimplantation, and death.
Related Concept Videos
Pharmacovigilance
This process, termed pharmacovigilance, aims to detect, evaluate, and minimize harmful effects related to medication use. The data collection for pharmacovigilance depends on spontaneous reporting systems, where healthcare professionals or patients voluntarily report suspected ADRs.
In some cases, there...
Mitral Stenosis III: Medical Management
Mitral Regurgitation III: Medical Management
Errors occurring during blood pressure monitoring
Several factors...
Mitral Valve Prolapse II: Assessment and Management
Aortic Regurgitation III: Medical Management