Low dose CTPA using a low kV technique combined with high IR: A clinical study
A Bellizzi1, P Bezzina1, F Zarb1
1Department of Radiography, Faculty of Health Sciences, University of Malta, Msida, Malta.
Introduction:
To investigate optimising a computerised tomography pulmonary angiogram (CTPA) scan protocol in terms of radiation dose and image quality using a low kV technique combined with high iterative reconstruction (IR) parameters (>50%) and apply the optimised protocol in clinical practice on patients irrespective of their body weight.
Methods:
CTPA examinations were performed on 64 patients equally divided into control and experimental groups. Patients in the control group were scanned using the current protocol (100 kV with 50% IR) while patients in the experimental group were scanned using an optimised protocol (80 kV with 60%IR). The radiation dose indices volume computerised tomography dose index (CTDIvol), dose length product (DLP), size specific dose estimates (SSDE) and effective dose (ED) were recorded. Subjective image quality was evaluated by 3 radiologists through absolute visual grading analysis (VGA) using an image quality scoring tool. The resultant image quality scores were analysed using Visual Grading Characteristics (VGC). Objective image quality was recorded in terms of contrast-to-noise-ratio (CNR) and signal-to-noise-ratio (SNR).
Results:
The application of the optimised protocol resulted in a statistically significant (p < 0.05) reduction in mean CTDIvol (-49%), DLP (-48%), SSDE (-52%) and ED (-49%). Objective image quality was significantly (p < 0.05) improved both in CNR (32%) and SNR (13%). Subjective image quality scores were higher for the current protocol but variation between the two protocols was not significant (p = 0.650).
Conclusions:
When applying the low kV technique combined with high IR parameters, a significant dose reduction may be achieved while still maintaining diagnostic image quality.
Implications For Practice:
The low kV technique combined with high IR parameters is an effective optimisation technique which can be easily implemented for the CTPA protocol.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Treatment for Pulmonary Arterial Hypertension: Receptor Tyrosine Kinase Inhibitors and Calcium Channel Blockers
TKIs, such as imatinib (Gleevec), are particularly effective in tackling the growth and mitogenic factors that become upregulated in PAH patients. These factors contribute to the...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography