Effects of contrast medium viscosity into flushing port on artefacts during optical coherence tomography imaging
Hiroki Okabe1, Konosuke Inoue1, Masamitsu Tanaka2
1The Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan.
Background:
Optical coherence tomography (OCT) is becoming the standard imaging modality for percutaneous coronary intervention (PCI) because of its high resolution. To perform appropriate OCT-guided PCI, it is necessary to avoid artefacts and obtain high-quality images. We investigated the relationship between artefacts and the viscosity of contrast media, which were used to remove air before OCT imaging catheter was inserted into guiding catheter.
Methods:
We retrospectively analyzed every pullback of OCT examinations from January 2020 to September 2021. Cases were divided into two groups according to the type of contrast media used for catheter flushing: low-viscosity (Iopamidol-300, Bayer, Nordrhein-Westfalen, Germany) vs. high-viscosity (Iopamidol-370, Bayer). We evaluated the artefacts and quality of each OCT image and performed ex vivo experiments to compare differences in artefact frequencies using the two contrast media.
Results:
A total of 140 pullbacks in the low-viscosity group and 73 pullbacks in the high-viscosity group were analyzed. The percentage of grade 2 and 3 images (with good quality) in the low-viscosity group was significantly lower (68.1 % vs. 94.5 %, p < 0.001). Rotational artefacts were significantly more common in the low-viscosity group (49.3 % vs. 8.2 %, p < 0.001). In multivariate analysis, using low-viscosity contrast media was a significant factor influencing the appearance of rotational artefacts and affecting image quality (odds ratio, 9.42; 95 % confidence interval, 3.58 to 24.8; p < 0.001). In ex vivo experiments, using low-viscosity contrast media was also a significant predictor of artefact occurrence during OCT (p < 0.01).
Conclusions:
The viscosity of the contrast agent used while flushing the OCT imaging catheter contributes to the appearance of OCT artefacts.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...