Related Experiment Video
Updated: Jul 29, 2025

Knowledge Based Cloud FE Simulation of Sheet Metal Forming Processes
Published on: December 13, 2016
Optimization of wear behavior of heat-treated Ti-6Al-7Nb biomedical alloy by response surface methodology
Syed Altaf Hussain1, Manoj Panchal1, K Viswanath Allamraju2
1Department of Mechanical Engineering, Rajeev Gandhi Memorial College of Engineering and Technology, Nandyal, 518501, India.
Abstract:
Titanium-based metals are used most often in biomedical implant studies because they have good qualities like being biocompatible, not being poisonous, Osseo-integration, high specific properties, wear resistance, etc. The main goal of this work is to improve the wear resistance of Ti-6Al-7Nb biomedical metal by using a mix of Taguchi, ANOVA, and Grey Relational Analysis. The effect of changeable control process factors like applied load, spinning speed, and time on wear reaction measures like wear rate (WR), coefficient of friction (COF), and frictional force. The optimal combinations of wear rate, COF, and frictional force minimise wear characteristics. The L9 Taguchi orthogonal array was used to plan the experiments, which were done on a pin-on-disc set-up according to ASTM G99. To find the best set of control factors, Taguchi, ANOVA, and Grey relationship analysis were used. The results show that a load of 30 N, a speed of 700 rpm, and a time of 10 min are the best control settings.