Related Experiment Video
Updated: Jul 29, 2025

Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Unusual atrial pacing induced by type II far-field p-wave sensing in a patient with complete atrioventricular block
Yasunobu Yamagishi1, Yasushi Oginosawa1, Nozomu Ishii1
1The Second Department of Internal Medicine, University of Occupational and Environmental Health, Kitakyushu, Japan.
Abstract:
An 88-year-old Japanese woman underwent DDD pacemaker (MicroPort KORA 250 DR, V lead: VEGA R52) implantation for complete atrioventricular block. A 12-lead electrocardiogram for a routine examination showed atrial pacing within the intrinsic P wave, followed by inhibition of ventricular pacing. Pacemaker interrogation revealed no abnormalities in the basic parameters; however, ventricular pacing was inhibited by far-field sensing of intrinsic atrial waves before atrial events; type II far-field P-wave sensing. As a result, unusual atrial pacing occurred due to the pause suppression algorithm, which is the one of the functions that prevent atrial fibrillation development.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Mechanism of Cardiac Arrhythmias