Development of phantom for current modulation quality assurance test on computed tomography.
Gustavo E de Camargo1, Glaube N Carneiro1, Jessica V Real2
1Graduate Program in Physics and Astronomy, Federal University of Technology-Parana, Curitiba, Brazil.
Radiation Protection Dosimetry
|May 24, 2023
Summary
A new phantom was developed for computed tomography (CT) automatic tube current modulation (ATCM) quality control. This tool effectively assesses how CT scanners adjust radiation dose based on patient size, ensuring safety and efficacy.
Area of Science:
- Medical Physics
- Radiological Technology
- Diagnostic Imaging
Background:
- Automatic tube current modulation (ATCM) is crucial for minimizing patient radiation dose in computed tomography (CT).
- Quality control (QC) for ATCM ensures CT systems accurately adapt radiation output to varying object sizes.
- Existing QC methods may not adequately assess ATCM performance across different attenuation levels.
Purpose of the Study:
- To develop and validate a dedicated phantom for assessing the performance of ATCM in CT scanners.
- To evaluate the capability of CT systems to modulate tube current in response to changes in object attenuation using the developed phantom.
- To ensure compliance with national and international quality assurance recommendations for CT examinations.
Main Methods:
- A novel phantom was constructed from high-density polyethylene, featuring a cylindrical shape with three distinct sizes.
- The phantom was utilized to conduct ATCM quality control tests on two different CT scanner models (Toshiba and Philips).
- Measurements focused on evaluating the CT system's adaptation of tube current as a function of the phantom's size and attenuation.
Main Results:
- The study demonstrated a clear correlation between discrete changes in phantom size and corresponding adjustments in tube current.
- CT scanners exhibited appropriate adaptation of tube current when presented with distinct shifts in object attenuation.
- The results confirm the CT system's ability to modulate radiation output effectively based on object characteristics.
Conclusions:
- The developed high-density polyethylene phantom is a suitable tool for performing ATCM quality control tests in CT.
- This phantom facilitates the verification of CT scanner performance in adapting radiation dose according to object size.
- The phantom supports adherence to quality assurance standards, enhancing patient safety during CT scans.
Related Concept Videos
Computed Tomography
4.6K
Tomography refers to imaging by sections. Computed tomography (CT) is a non-invasive imaging technique that uses computers to analyze several cross-sectional X-rays to reveal minute details about structures in the body.
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
4.6K
Imaging Studies III: Computed Tomography
30
DefinitionComputed Tomography (CT) of the genitourinary (GU) tract is a non-invasive imaging modality that utilizes X-rays and computer processing to generate detailed cross-sectional images of the urinary system, encompassing the kidneys, ureters, bladder, and adjacent structures such as the adrenal glands.PurposeCT scans of the GU tract serve several diagnostic and therapeutic purposes, including:Diagnosis of Urinary Tract Diseases: Detects kidney stones, tumors, cysts, and congenital...
30