Related Experiment Video
Updated: Jul 31, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Simple one-parameter function to retrieve the correct exposure value from CR-39 radon detectors in high-saturation
Rocco Marchese1, Antonella Nicolino2, Achille Palma1
1ARPAB, Agenzia Regionale per la Protezione dell'Ambiente di Basilicata, 85100 - Potenza, Italy.
Abstract:
Radon activity concentration measurements in air can be easily performed by using CR-39 detectors whose response is almost linear in the range of medium-low exposures. However, when the exposure values grow too much, a saturation phenomenon becomes relevant and some corrections have to be adopted even if they cannot always be very accurate and easy to apply. Therefore, a simple alternative method for determining the correct response curve of the CR-39 detectors, from low up to very high radon exposures, is showed. To verify its robustness and the general applicability, several certified measurements were carried out in a radon chamber at different levels of exposure. Moreover, two different types of commercially available radon analysis systems were used.
More Related Videos