Case Series of Ventricular Tachycardia Ablation With Pulsed-Field Ablation: Pushing Technology Further (Into the
Cristina Lozano-Granero1, Jun Hirokami2, Eduardo Franco1
1University Hospital Ramón y Cajal, Madrid, Spain; Centro de Investigación Biomédica en Red de Enfermedades Cardiovasculares, Madrid, Spain.
Abstract:
Three cases of ventricular tachycardia ablation with pulsed-field ablation technology performed at 2 separate centers are reported, highlighting the advantages and disadvantages of this tool inside the ventricle: its dependence on proximity rather than contact makes it useful in sites with poor stability, while the speed of application and large scope of action provided by commercially available catheters could help with ablating large diseased areas of endocardium in a fast and hemodynamically well-tolerated fashion. However, lesion depth could be insufficient for guaranteeing efficacy in preventing ventricular tachycardias originating at an epicardial site, even in the right ventricle.