Related Experiment Video
Updated: Jul 29, 2025

Author Spotlight: Investigating Islet Abnormalities and Function with a Pseudoislet Protocol
Published on: November 3, 2023
In silico design and validation of a time-varying PID controller for an artificial pancreas with intraperitoneal
Alberto Dalla Libera1, Chiara Toffanin2, Martina Drecogna1
1Department of Woman and Child's Health, University of Padova, 35128 Padova, Italy.
Abstract:
Type 1 diabetes (T1D) is a chronic autoimmune disease featured by the loss of beta cell function and the need for lifetime insulin replacement. Over the recent decade, the use of automated insulin delivery systems (AID) has shifted the paradigm of treatment: the availability of continuous subcutaneous (SC) glucose sensors to guide SC insulin delivery through a control algorithm has allowed, for the first time, to reduce the daily burden of the disease as well as to abate the risk for hypoglycemia. AID use is still limited by individual acceptance, local availability, coverage, and expertise. A major drawback of SC insulin delivery is the need for meal announcement and the peripheral hyperinsulinemia that, over time, contributes to macrovascular complications. Inpatient trials using intraperitoneal (IP) insulin pumps have demonstrated that glycemic control can be improved without meal announcement due to the faster insulin delivery through the peritoneal space. This calls for novel control algorithms able to account for the specificities of IP insulin kinetics. Recently, our group described a two-compartment model of IP insulin kinetics demonstrating that the peritoneal space acts as a virtual compartment and IP insulin delivery is virtually intraportal (intrahepatic), thus closely mimicking the physiology of insulin secretion. The FDA-accepted T1D simulator for SC insulin delivery and sensing has been updated for IP insulin delivery and sensing. Herein, we design and validate-in silico-a time-varying proportional integrative derivative controller to guide IP insulin delivery in a fully closed-loop mode without meal announcement.
More Related Videos
07:55A Multi-Parametric Islet Perifusion System within a Microfluidic Perifusion Device
Published on: January 26, 2010
09:08Three-dimensional Printing of Thermoplastic Materials to Create Automated Syringe Pumps with Feedback Control for Microfluidic Applications
Published on: August 30, 2018
Related Concept Videos
PID Controller
PI Controller: Design
Time and frequency -Domain Interpretation of PI Control
Acting as a low-pass filter, the PI controller slows the system's response and extends settling times. This requires...
PD Controller: Design
Designing a continuous-data controller requires selecting and linking components like adders and integrators, which are fundamental in Proportional,...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...