Rapid Multiplex RT-PCR for Influenza A and B by Genesoc®, a Microfluidic PCR System
Miyako Takata1, Masaki Nakamoto2, Tsuyoshi Kitaura2
1Department of Pathobiological Science and Technology, Graduate School of Medical Science, School of Health Science, Faculty of Medicine, Tottori University, Yonago 683-8503, Japan.
Yonago Acta Medica
|May 25, 2023
Summary
A new microfluidic real-time PCR system, GeneSoC®, offers a rapid and sensitive method for diagnosing influenza A and B. This molecular test shows high accuracy, making it a promising alternative to less sensitive rapid antigen tests.
Area of Science:
- Molecular Diagnostics
- Virology
- Microfluidics
Background:
- Rapid antigen tests are common for influenza diagnosis but have low sensitivity.
- There is a need for more sensitive molecular diagnostic tests for influenza A and B.
- Microfluidic thermal cycling technology offers potential for rapid molecular testing.
Purpose of the Study:
- To develop and clinically evaluate a rapid multiplex testing protocol for influenza A and B using the GeneSoC® microfluidic real-time PCR system.
- To compare the performance of the GeneSoC® assay against conventional real-time RT-PCR and rapid antigen tests.
Main Methods:
- Developed a rapid real-time PCR protocol on the GeneSoC® microfluidic system for influenza A and B detection.
- Validated specificity with influenza A/B, human metapneumovirus, and RSV strains.
- Evaluated analytical sensitivity using synthesized RNA and clinical nasopharyngeal samples.
- Performed cross-validation against conventional RT-PCR and rapid antigen tests.
Main Results:
- The GeneSoC® assay detected influenza A and B at low concentrations (38 and 65 copies/µL).
- 100% agreement was observed between GeneSoC® RT-PCR and conventional RT-PCR for clinical specimens.
- GeneSoC® RT-PCR showed 95.7% overall agreement with rapid antigen tests.
- The average turnaround time for GeneSoC® RT-PCR was approximately 16.5 minutes.
Conclusions:
- The GeneSoC® microfluidic real-time PCR system demonstrates analytical performance comparable to conventional RT-PCR.
- The system offers a rapid turnover time, making it a viable alternative to current rapid antigen tests for influenza diagnosis.