Related Experiment Video
Updated: Jul 29, 2025

Measuring Pressure Volume Loops in the Mouse
Published on: May 2, 2016
Comparison of Mechanical Forces used in Open Tracheotomy versus Percutaneous Tracheotomy Techniques
Ramya Bharathi1, Gopikrishna M Rao2, Jeremiah Tracy1,2
1Department of Otolaryngology, Tufts Medical Center, Boston, Massachusetts, USA.
Objective:
To understand the etiology of tracheotomy-induced tracheal stenosis by comparing the differences in techniques and mechanical force applied with open tracheotomy (OT) versus percutaneous tracheotomy (PCT) placement.
Methods:
This study is an unblinded, experimental, randomized controlled study in an ex-vivo animal model. Simulated tracheostomies were performed on 10 porcine tracheas, 5 via a tracheal window technique (OT) and 5 using the Ciaglia technique (PCT). The applied weight during the simulated tracheostomy and the compression of the trachea were recorded at set times during the procedure. The applied weight during tracheostomy was used to calculate the tissue force in Newtons. Tracheal compression was measured by anterior-posterior distance compression and as percent change.
Results:
Average forces for scalpel (OT) versus trocar (PCT) were 2.6 N and 12.5 N (p < 0.01), with the dilator (PCT) it was 22.02 N (p < 0.01). The tracheostomy placement with OT required an average force of 10.7 N versus 23.2 N (p < 0.01) with PCT. The average change in AP distance when using the scalpel versus trocar was 21%, and 44% (p < 0.01), with the dilator it was 75% (p < 0.01). The trach placement with OT versus PCT had an average AP distance change of 51% and 83% respectively (p < 0.01).
Conclusion:
This study demonstrated that PCT required more force and caused more tracheal lumen compression when compared to the OT technique. Based on the increased force required for PCT, we suspect there could also be an increased risk for tracheal cartilage trauma.
Level Of Evidence:
NA Laryngoscope, 134:103-107, 2024.
More Related Videos
04:43Endotracheal Intubation via Tracheotomy and Subsequent Thoracotomy in Rats for Non-Survival Applications
Published on: March 15, 2024
09:10Heterotopic and Orthotopic Tracheal Transplantation in Mice used as Models to Study the Development of Obliterative Airway Disease
Published on: January 20, 2010
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...