Impacts of different reconstruction methods on the image quality of cadmium-zinc-telluride-based single photon
Weiming Wu1, Ruyi Zhang1, Yaqian Zhou2
1Department of Nuclear Medicine, Tianjin Medical University General Hospital, Tianjin.
Objective:
The objective was to evaluate the impacts of different reconstruction methods [filtered back projection (FBP) and ordered subset expectation maximization (OSEM)] and different filters (Butterworth filter and Gaussian filter) on the image quality in cadmium-zinc-telluride (CZT)-based single photon emission computed tomography (SPECT)/computed tomography (CT) pulmonary perfusion imaging.
Methods:
A combinations including FBP with Butterworth filter, OSEM with Butterworth filter (OSEM + Butterworth filter ), and OSEM with Gaussian filter (OSEM + Gaussian filter) were used during SPECT image reconstruction. Visual and quantitative parameters [root mean square (RMS) noise, contrast and contrast-to-noise ratio (CNR)] were used to evaluate image quality.
Results:
The OSEM + Gaussian filter had better RMS noise and CNR than those of the FBP + Butterworth filter or OSEM + Butterworth filter, while the OSEM + Butterworth filter had the best contrast. The highest visual scores were obtained by OSEM + Gaussian filter ( P < 0.0001). In the lesion size <2 cm group, the contrast ( P < 0.01) and visual scores ( P < 0.001) of OSEM + Butterworth filter were better than those of the other two groups. In the lesion size ≥2 cm group, the RMS noise and visual scores of OSEM + Gaussian filter were better than those of the other two groups.
Conclusion:
In CZT SPECT/CT pulmonary perfusion imaging, this study recommended the clinical use of the OSEM + Gaussian filter combination for reconstruction in both conventional and larger lesions, the OSEM + Butterworth filter image postprocessing method might be advantageous in small lesions.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...