Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Accuracy of a Standalone Atrial Fibrillation Detection Algorithm Added to a Popular Wristband and Smartwatch:
Jasper L Selder1, Henryk Jan Te Kolste1, Jos Twisk2
1Department of Cardiology, Amsterdam University Medical Center, Amsterdam, Netherlands.
Background:
Silent paroxysmal atrial fibrillation (AF) may be difficult to diagnose, and AF burden is hard to establish. In contrast to conventional diagnostic devices, photoplethysmography (PPG)-driven smartwatches or wristbands allow for long-term continuous heart rhythm assessment. However, most smartwatches lack an integrated PPG-AF algorithm. Adding a standalone PPG-AF algorithm to these wrist devices might open new possibilities for AF screening and burden assessment.
Objective:
The aim of this study was to assess the accuracy of a well-known standalone PPG-AF detection algorithm added to a popular wristband and smartwatch, with regard to discriminating AF and sinus rhythm, in a group of patients with AF before and after cardioversion (CV).
Methods:
Consecutive consenting patients with AF admitted for CV in a large academic hospital in Amsterdam, the Netherlands, were asked to wear a Biostrap wristband or Fitbit Ionic smartwatch with Fibricheck algorithm add-on surrounding the procedure. A set of 1-min PPG measurements and 12-lead reference electrocardiograms was obtained before and after CV. Rhythm assessment by the PPG device-software combination was compared with the 12-lead electrocardiogram.
Results:
A total of 78 patients were included in the Biostrap-Fibricheck cohort (156 measurement sets) and 73 patients in the Fitbit-Fibricheck cohort (143 measurement sets). Of the measurement sets, 19/156 (12%) and 7/143 (5%), respectively, were not classifiable by the PPG algorithm due to bad quality. The diagnostic performance in terms of sensitivity, specificity, positive predictive value, negative predictive value, and accuracy was 98%, 96%, 96%, 99%, 97%, and 97%, 100%, 100%, 97%, and 99%, respectively, at an AF prevalence of ~50%.
Conclusions:
This study demonstrates that the addition of a well-known standalone PPG-AF detection algorithm to a popular PPG smartwatch and wristband without integrated algorithm yields a high accuracy for the detection of AF, with an acceptable unclassifiable rate, in a semicontrolled environment.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
15:00Setup of Consumer Wearable Devices for Exposure and Health Monitoring in Population Studies
Published on: February 3, 2023
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias
Assessment of blood pressure in brachial artery(two-step method)
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation