Measurement of Factor V Activity in Human Plasma Using a Microplate Coagulation Assay
Published on: September 9, 2012
Sensitive Measurement of Clinically Relevant Factor VIII Levels in Thrombin Generation Assays Requires Presence of
Tom W van de Berg1,2, Erik A M Beckers2, Floor C J I Heubel-Moenen2
1Department of Biochemistry, Cardiovascular Research Institute Maastricht (CARIM), University Maastricht, Maastricht, The Netherlands.
Background:
Hemophilia A (HA) is characterized by decreased or absent factor VIII (FVIII) activity. Current FVIII assays are based on clotting time and thus only provide information about the initiation of coagulation. In contrast, thrombin generation assays (TGAs) can be used to measure the full coagulation spectrum of initiation, propagation, and termination that provide information on the whole course of thrombin generation and inhibition. However, the commercially available TG kits lack sensitivity for measurements of hemophilia plasma within lower FVIII ranges, which is essential for explaining differences in bleeding phenotypes in hemophiliacs at clinically low levels of FVIII.
Aims:
Optimization of the TGA for measurements of low FVIII levels in severe HA patients.
Methods:
TGA measurements were performed in severe HA pooled plasma (n = 10). Investigations of several preanalytical and analytical variables of the assay were performed in a stepwise process and adjusted based on sensitivity toward intrinsic coagulation activation.
Results:
TGA initiated by tissue factor (TF) alone at varying concentrations was unable to significantly differentiate between FVIII levels below 20%. In contrast, TGA activation with low concentrations of TF in presence of FXIa appeared to be highly sensitive for FVIII changes both in high and low ranges. In addition, a representative TGA curve at trough levels could only be produced using the dual TF/FXIa TGA.
Conclusion:
We propose a critical optimization for the setup of the TGA for measurements in severe HA plasma. The dual TF/FXIa TGA shows increased sensitivity, especially in lower FVIII ranges, which allows for better individual characterization at baseline, prediction of interventions, and follow-up.
More Related Videos
03:53Author Spotlight: High-Sensitivity Tissue Factor Activity Assay for Plasma Diagnosis
Published on: December 29, 2023
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
Related Concept Videos
Venous Thrombosis III: Interprofessional Care
Anticoagulant Drugs: Low-Molecular-Weight Heparins