CT Imaging of Eustachian Tube Balloon Dilation: Method Development on Cadaver Heads
Selma Cetin-Ferra1, Miriam S Teixeira2, J Douglas Swarts3
1Department of Cell Biology, University of Pittsburgh School of Medicine, Pittsburgh, PA 15260, USA.
Bioengineering (Basel, Switzerland)
|May 27, 2023
Summary
This study introduces a new method using CT scans to measure changes in the Eustachian tube (ET) after balloon dilation (BD). The technique quantifies the effectiveness of BD by analyzing air density changes within the ET lumen.
Area of Science:
- Otolaryngology
- Medical Imaging
- Anatomy
Background:
- The Eustachian tube (ET) plays a crucial role in middle ear pressure regulation.
- Balloon dilation (BD) is an emerging treatment for ET dysfunction.
- Quantifying the structural effects of BD on the ET is essential for evaluating treatment efficacy.
Purpose of the Study:
- To develop and validate a novel methodology for assessing the impact of balloon dilation (BD) on Eustachian tube (ET) structure.
- To utilize Computerized Tomography (CT) imaging for quantitative analysis of ET lumen changes post-BD.
- To establish a reliable method for measuring the success of ET balloon dilation.
Main Methods:
- BD was performed on human cadaver ETs.
- Axial CT images were acquired before, during, and after balloon dilation.
- ImageJ software and 3D volume viewer were used to match anatomical landmarks and capture serial ET images.
- Region of interest (ROI) histograms were analyzed to determine air density changes, serving as a measure of dilation.
Main Results:
- A small ROI effectively captured the most significant ET lumen dilation post-BD.
- Average air density increase in the small ROI was 64%, indicating successful dilation.
- Longer and longest ROI measurements showed air density increases of 56% and 44%, respectively.
Conclusions:
- A reproducible CT-based method was developed to image and quantify ET changes after BD.
- The methodology allows for objective measurement of BD outcomes using anatomical landmarks and air density analysis.
- This approach provides a valuable tool for evaluating the efficacy of Eustachian tube balloon dilation.