Clinical Feasibility of Deep Learning-Based Image Reconstruction on Coronary Computed Tomography Angiography
Seul Ah Koo1, Yunsub Jung2, Kyoung A Um2
1Department of Radiology and The Research Institute of Radiological Science, Gangnam Severance Hospital, Yonsei University College of Medicine, Seoul 06273, Republic of Korea.
Abstract:
This study evaluated the feasibility of deep-learning-based image reconstruction (DLIR) on coronary computed tomography angiography (CCTA). By using a 20 cm water phantom, the noise reduction ratio and noise power spectrum were evaluated according to the different reconstruction methods. Then 46 patients who underwent CCTA were retrospectively enrolled. CCTA was performed using the 16 cm coverage axial volume scan technique. All CT images were reconstructed using filtered back projection (FBP); three model-based iterative reconstructions (MBIR) of 40%, 60%, and 80%; and three DLIR algorithms: low (L), medium (M), and high (H). Quantitative and qualitative image qualities of CCTA were compared according to the reconstruction methods. In the phantom study, the noise reduction ratios of MBIR-40%, MBIR-60%, MBIR-80%, DLIR-L, DLIR-M, and DLIR-H were 26.7 ± 0.2%, 39.5 ± 0.5%, 51.7 ± 0.4%, 33.1 ± 0.8%, 43.2 ± 0.8%, and 53.5 ± 0.1%, respectively. The pattern of the noise power spectrum of the DLIR images was more similar to FBP images than MBIR images. In a CCTA study, CCTA yielded a significantly lower noise index with DLIR-H reconstruction than with the other reconstruction methods. DLIR-H showed a higher SNR and CNR than MBIR (p < 0.05). The qualitative image quality of CCTA with DLIR-H was significantly higher than that of MBIR-80% or FBP. The DLIR algorithm was feasible and yielded a better image quality than the FBP or MBIR algorithms on CCTA.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography