Measuring the Motor Aspect of Cancer-Related Fatigue using a Handheld Dynamometer
Published on: February 20, 2020
Reliability and Validity of Shoulder and Handgrip Strength Testing
Nicholas Robert Biasini1, Brian Bannon1, Michael Pellegrino1
1Department of Physical Therapy, University of Toronto, Toronto, Ontario, Canada.
The JTECH wireless device shows high reliability for measuring shoulder and handgrip strength. It also demonstrates strong validity when compared to established strength testing equipment.
Area of Science:
- Biomechanics
- Musculoskeletal assessment
- Rehabilitation technology
Background:
- Accurate measurement of maximal isometric strength is crucial for assessing shoulder and hand function.
- Existing strength testing devices vary in reliability and validity.
- The JTECH computerized, wireless apparatus offers a novel approach to strength assessment.
Purpose of the Study:
- To evaluate the intra- and inter-rater reliability of the JTECH computerized, wireless apparatus for shoulder isometric and handgrip strength measurements.
- To determine the validity of the JTECH device by comparing its measurements to established dynamometers (Micro-FET2 and Jamar).
- To assess the JTECH device's performance in healthy adults without shoulder pathology.
Main Methods:
- Twenty healthy young adults participated in the study.
- Shoulder isometric strength was measured using JTECH and Micro-FET2 dynamometers.
- Handgrip strength was measured using JTECH and Jamar dynamometers.
- Intra-rater reliability and convergent validity were assessed by one rater over two days.
- Inter-rater reliability was assessed by a second rater on a third visit.
Main Results:
- The JTECH device exhibited excellent intra-rater reliability (ICC = 0.78-0.97) and inter-rater reliability (ICC = 0.76-0.95) for strength measures.
- Substantial concurrent validity was found for shoulder flexion (R=0.87), extension (R=0.87), abduction (R=0.88), and adduction (R=0.85) when compared to the Micro-FET2.
- Substantial concurrent validity was also demonstrated for handgrip strength (R=0.92) between the JTECH and Jamar devices.
Conclusions:
- The JTECH computerized, wireless devices are reliable for both intra- and inter-rater measurements of shoulder isometric and handgrip strength.
- The JTECH device demonstrates substantial concurrent validity against established strength-measuring tools.
- These findings support the use of the JTECH wireless apparatus for accurate strength assessments in clinical and research settings.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:44Manual Muscle Testing: A Method of Measuring Extremity Muscle Strength Applied to Critically Ill Patients
Published on: April 12, 2011
Related Concept Videos
Reliability and Validity
Pre-Procedural Guidelines for Assessing Blood Pressure