Related Experiment Video
Updated: Jul 28, 2025

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
Direct visualization of proton beam irradiation effects in liquids by MRI
Sebastian Gantz1,2, Leonhard Karsch1,2, Jörg Pawelke1,2
1OncoRay-National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden-Rossendorf, 01309 Dresden, Germany.
Abstract:
The main advantage proton beams offer over photon beams in radiation therapy of cancer patients is the dose maximum at their finite range, yielding a reduction in the dose deposited in healthy tissues surrounding the tumor. Since no direct method exists to measure the beam's range during dose delivery, safety margins around the tumor are applied, compromising the dose conformality and reducing the targeting accuracy. Here, we demonstrate that online MRI can visualize the proton beam and reveal its range during irradiation of liquid-filled phantoms. A clear dependence on beam energy and current was found. These results stimulate research into novel MRI-detectable beam signatures and already find application in the geometric quality assurance for magnetic resonance-integrated proton therapy systems currently under development.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging