Related Experiment Video
Updated: Jul 28, 2025

Treatment of Facial Deformities using 3D Planning and Printing of Patient-Specific Implants
Published on: May 23, 2020
A Novel Orthognathic Surgery With a Half-Millimeter Accuracy for the Maxillary Positioning Using Prebent Plates and
Yosuke Yamashita1, Haruki Imai1, Hikaru Takasu1
1Department of Oral and Maxillofacial Surgery/Orthodontics, Yokohama City University Medical Center.
Abstract:
This retrospective study aimed to assess the accuracy of prebent plates and computer-aided design and manufacturing osteotomy guide for orthognathic surgery. The prebent plates correspondent to the planning model were scanned with a 3-dimensional printed model for guide design and used for fixation. Forty-two patients who underwent bimaxillary orthognathic surgery using computer-aided design and manufacturing intermediate splint with the guide (guided group: 20 patients) or with conventional fixation under straight locking miniplates (SLMs) technique (SLM group: 20 patients) were analyzed. A deviation of the maxilla between the planned and postoperative positions was evaluated using computed tomography, which was taken 2 weeks before and 4 days after the surgery. The surgery time and the infraorbital nerve paranesthesia were also evaluated. The mean deviations in the mediolateral ( x ), anteroposterior ( y ), and vertical directions ( z ) were 0.25, 0.50, and 0.37 mm, respectively, in the guided group, while that in the SLM group were 0.57, 0.52, and 0.82 mm, respectively. There were significant differences in x and z coordinates ( P <0.001). No significant difference in the surgery duration and paranesthesia was seen, suggesting the present method offers a half-millimeter accuracy for the maxillary repositioning without increasing the risk of extending surgery duration and nerve complication.