Related Experiment Video
Updated: Jul 28, 2025

Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
Enhanced Aerosol Delivery During High-Flow Nasal Cannula Therapy
Jane Moon1, Michael McPeck2, Jeyanthan Jayakumaran2
1Pulmonary, Critical Care and Sleep Medicine Division, Department of Medicine, Stony Brook University Medical Center, Stony Brook, New York. Jane.moon@stonybrookmedicine.edu.
Background:
Aerosolized drug delivery via high-flow nasal cannula (HFNC) decreases as gas flow is increased. To improve aerosol delivery, breath-enhanced jet nebulizer may increase aerosol output. This study tested that hypothesis and compared breath-enhanced jet nebulizer to vibrating mesh nebulizer technology.
Methods:
First, in an isolated circuit, breath-enhanced jet nebulizer and vibrating mesh nebulizer aerosol outputs were measured during simulated HFNC by using infused saline solution at rates of 5-60 mL/h. Limits were defined when nebulizer filling was detected. The devices were then tested by using 99mTc/saline solution to measure maximum rates of aerosol production. After the output experiments, drug delivery was measured in vitro by using a model that consisted of an HFNC circuit interfaced to a realistic 3-dimensional printed head. The 99mTc/saline solution was infused at rates of 5 to 60 mL/h for the breath-enhanced jet nebulizer and 5 to 20 mL/h for the vibrating mesh nebulizer with HFNC gas flows of 10 to 60 L/min. Aerosol delivery to the trachea was measured by using a shielded ratemeter, which defined the rate of drug delivery (µg NaCl/min).
Results:
With increasing gas flow, breath-enhanced jet nebulizer output increased to a maximum of 50 mL/h, the vibrating mesh nebulizer maximum was 12 mL/h. At HFNC gas flow of 60 L/min, breath-enhanced jet nebulizer delivered 3.16 to 316.8 µg NaCl/min, the vibrating mesh nebulizer delivered 23.5 to 61.7 µg NaCl/min. For infusion pump flows of 5 to 12 mL/h, the rate of drug delivery was independent of nebulizer type (P = .19) and dependent on infusion pump flow (P < .001) and gas flow (P < .001).
Conclusions:
Increasing gas flow increased breath-enhanced jet nebulizer output, which demonstrated the effects of breath enhancement. At 60 L/min, breath enhanced jet nebulizer delivered up to 5 times more aerosol compared with conventional vibrating mesh nebulizer technology. Breath-enhanced jet nebulizer delivered a wide range of dose rates at all high flows. In patients who are critically ill, breath-enhanced jet nebulizer technology may allow titration of bedside dosing based on clinical response by simple adjustment of the infusion rate.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
05:43Topical Airway Anesthesia for Awake-endoscopic Intubation Using the Spray-as-you-go Technique with High Oxygen Flow
Published on: January 13, 2017
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Inhaled Medications
Suctioning the Nasopharyngeal Airway
Equipment Required
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: