Dosimetric effect of six degrees of freedom couch top with rotational setup error corrections in proton therapy
Takahiro Kato1,2, Sho Sasaki2, Tomohiro Ikeda2
1Department of Radiological Sciences, School of Health Sciences, Fukushima Medical University, Fukushima, Japan.
Purpose:
To investigate the dosimetric effect of six degrees of freedom (6DoF) couch top with rotational corrections in proton therapy (PT).
Methods:
The water equivalent thickness (WET) was measured using a proton beam with a 6DoF couch top and patient immobilization base plate (PIBP) placed in front of a motorized water phantom. The accuracy verification was performed with the beam axis set perpendicular to the 6DoF couch top and tilted in 10° steps from 10° to 30°. Up to 3° rotational correction may be added during the actual treatment to correct the rotational setup error on our system. The measured and calculated values using the treatment planning system were compared. Additionally, the effect of the 3° difference was evaluated using actual measurements concerning each angle on the proton beam range.
Results:
The WET of the 6DoF couch top and PIBP were 8.5 ± 0.1 mm and 6.8 ± 0.1 mm, respectively. The calculation and the actual measurement at each angle agreed within 0.2 mm at the maximum. A maximum difference of approximately 0.6 mm was confirmed when tilted at 3° following 30° with the 6DoF couch top plus PIBP.
Conclusions:
The dosimetric effect of the 6DoF couch top with rotational corrections in PT differs depending on the incidence angle on the couch top, and it increased with the increased oblique angle of incidence. However, the effect on the range was as small as 0.6 mm at the maximum. The amount of rotational correction, the angle of incidence of the beam, and the effect of rotational corrections on the proton beam range may differ depending on the structure of the couch top. Therefore, sufficient prior confirmation, and subsequent periodical quality assurance management are important.
Related Concept Videos
One-Degree-of-Freedom System
A one-degree-of-freedom system is defined by an independent variable that determines its state and behavior. One example of a one-degree-of-freedom system is a simple harmonic oscillator, such as a...
Kinematic Equations for Rotation
For instance, imagine a point A on a rigid body engaged in circular motion. The translational velocity of this particular point can be calculated by taking the time derivatives of the displacement equation, which essentially measures the...
Relative Motion Analysis using Rotating Axes-Problem Solving
Here, in order to determine the magnitude of velocity and acceleration for point...