Hydrophilic interaction liquid chromatography with methanol-water eluent on a zeolite
James Otis1, Jörg Hofmann2, Thomas Schmidt3
1University of Toronto, Department of Biochemistry, 1 King's College Circle, Toronto, ONM5G 0A4, Canada.
Background:
Hydrophilic interaction chromatography (HILIC) works with organic solvent-water mixtures as eluent and is based on the formation of a water enriched liquid phase on the surface of a hydrophilic stationary phase. Hydrophilic solutes are retained on that stagnant water-rich film depending on the difference of solvation compared to the mobile phase composition. However, the enhancement of selectivity by increasing the fraction of organic cosolvent is coupled with a limitation the analyte solubility, and the improvement of the HILIC principle by new hydrophilic stationary phases is the remaining option.
Results:
Y-zeolite (faujasite, FAU type) in the Na+-form with an average particle diameter of 5 μm was used as packing material in a 125 mm long HPLC column. The chromatographic response of the column was tested in methanol-water mixtures as eluent after injection of several aliphatic alcohols, polyols and monosaccharides with eluent conditions where no separation occurs on diol functionalized silica. On the zeolite the retention time increases according to ethylene glycol < glycerol < erythritol < sorbitol < inositol. The separation principle is explained to be superposed by two effects: firstly, a partition equilibrium between the water-rich phase in the zeolite micropores exists, and secondly, selective interactions with the inner crystalline pore surface and fixed-position Na+ ions, both serving to enhance the selectivity. Furthermore, arabinose and fructose monosaccharides could be separated into their tautomeric forms. Only upon increasing the temperature from 20 to 60 °C the tautomeric pattern merges into a single peak.
Significance And Novelty:
Instead of the stagnant water rich surface layer, zeolite micropores now take over that function. As a result, the selectivity among polyols and between α/β-arabinopyranose and β-fructopyranose/β-fructofuranose tautomers is extraordinary superior towards conventional hydrophilic interaction liquid chromatography (HILIC).
More Related Videos
10:21Automated Hydrophobic Interaction Chromatography Column Selection for Use in Protein Purification
Published on: September 21, 2011
04:39A Modified QuEChERS-HPLC Method for Detection of Polycyclic Aromatic Hydrocarbons in Zebrafish Embryos Exposed to Fine Particulate Matter
Published on: June 13, 2025
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Thin-Layer Chromatography (TLC): Overview
To begin the analysis, a mixture of compounds is spotted on the starting line on the TLC plate using a thin capillary. The bottom of the...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...