Assay discrepancies using human coagulation factor VIII chromogenic kits: Results from a plasma-derived factor VIII
1National Institute for Biological Standards and Control (NIBSC), South Mimms, Potters Bar, Hertfordshire, EN6 3QG, United Kingdom.
Pharmeuropa Bio & Scientific Notes
|June 5, 2023
Summary
Chromogenic assay discrepancies in FVIII-product testing were confirmed between kits. Using observed, not prescribed, incubation times reduced potency differences, highlighting the need for lab-specific calibration.
Area of Science:
- Biotechnology and Pharmaceutical Analysis
- Hematology and Coagulation Diagnostics
Background:
- Discrepancies in chromogenic assays for Factor VIII (FVIII) products were reported by European laboratories.
- These variations impact the accurate potency determination of FVIII therapeutic concentrates.
Purpose of the Study:
- To investigate and identify the causes of FVIII chromogenic assay discrepancies.
- To compare the performance of different chromogenic kits (Coatest®, Coamatic®, Siemens) under controlled conditions.
Main Methods:
- A collaborative study involving multiple laboratories assessing FVIII products.
- Measurement of observed Factor X (FX) activation times (T1/2) for each kit.
- Comparison of observed T1/2 with manufacturer-prescribed T1/2 for potency calculations.
Main Results:
- Significant potency discrepancies were found between Coatest®/Coamatic® and Siemens/Coamatic® kits when using prescribed incubation times.
- Coamatic® kits generally yielded higher potencies.
- Observed T1/2 values varied significantly between laboratories and did not match prescribed times, impacting potency results.
Conclusions:
- Chromogenic FVIII potency assays can exhibit significant inter-kit and inter-laboratory variability.
- Using actual observed T1/2 incubation times, rather than prescribed ones, can reduce potency discrepancies.
- Laboratories should determine kit-specific FX activation curves to establish optimal T1/2 for accurate FVIII potency determination.