Ultrasound Imaging of Hip Displacement in Children With Cerebral Palsy

Thanh-Tu Pham1, Lawrence H Le1, Thanh-Giang La2

  • 1Department of Radiology and Diagnostic Imaging, University of Alberta, Edmonton, Alberta, Canada; Department of Biomedical Engineering, University of Alberta, Edmonton, Alberta, Canada.

Insights

This study presents a new ultrasound (US) method to estimate hip displacement in children. The developed migration percentage (MP) index shows accurate results validated by simulations and phantom studies.

Area of Science:

  • Medical imaging
  • Biomedical engineering
  • Pediatric orthopedics

Background:

  • Hip displacement in children requires accurate diagnostic methods.
  • Ultrasound (US) imaging offers a non-invasive approach for hip assessment.

Purpose of the Study:

  • To describe and validate an ultrasound-based method for estimating hip displacement.
  • To introduce a new diagnostic index, the migration percentage (MP).

Main Methods:

  • Developed a migration percentage (MP) index using acetabulum-femoral head distance and femoral head width from US images.
  • Validated the method using numerical simulations, 3-D-printed hip phantoms, and pilot in vivo data.

Main Results:

  • Numerical simulations showed a maximum diameter error of 16.1 ± 8.5% under specific noise and roughness conditions.
  • Phantom studies demonstrated low percentage errors for MP (0.3%-6.6% vs. X-ray).
  • Pilot clinical trials showed a mean absolute difference of 3.5 ± 2.8% between X-ray and US MPs.

Conclusions:

  • The developed ultrasound method is a viable tool for evaluating hip displacement in pediatric patients.
  • The migration percentage (MP) index provides accurate and reliable measurements.